Views: 0 Author: Site Editor Publish Time: 2026-11-16 Origin: Site
In blast shock wave measurement, pipeline transient pressure testing, fluid impact experiments and other test scenarios, pressure transients evolve on microsecond timescales. Conventional pressure sensors lack sufficient response speed and fail to capture critical peak pressure and rising‑edge information. The Y1004E‑4 piezoelectric pressure sensor is designed as a performance equivalent to comparable PCB sensors. Utilizing the piezoelectric effect, it serves as the core sensing element for high‑frequency transient pressure measurement.
1. Core Sensing Element: Quartz versus Piezoceramic
The Y1004E‑4 adopts single‑crystal piezoelectric quartz as its sensitive element. Benefiting from the intrinsic superior piezoelectric properties of quartz, it is an optimal solution for high‑frequency transient pressure testing.
As a single‑crystal piezoelectric material, quartz differs fundamentally from piezoceramics. It is free from poling degradation and Curie temperature limitations. It features stable charge output, ultra‑low thermal drift, high insulation resistance and minimal charge leakage, and exhibits excellent measurement repeatability over extended cyclic tests. This characteristic makes it well suited for blast shock wave and pulsed pressure tests that demand repeatable comparison and reproducible datasets.
The extremely high elastic stiffness of quartz crystal underpins its high‑frequency response capability. Subjected to shock loading, the crystal generates charge instantaneously via the piezoelectric effect, delivering ultra‑fast response tailored for rapidly varying dynamic pressure signals such as pressure pulses and shock waves.
Note: Piezoelectric quartz sensors are intended exclusively for dynamic pressure measurement. Charge dissipates gradually over time, so they cannot sustain static pressure measurement for prolonged durations. They are only applicable to alternating transient processes including pulses, shock waves and pulsating pressure.
Natural resonant frequency is one of the primary performance metrics for dynamic pressure sensors. The Y1004E‑4 specifies a natural resonant frequency ≥ 250 kHz.
A sensor constitutes a mechanical oscillatory system. When excited by a step pressure input, its mechanical structure undergoes inherent oscillation. Standard engineering convention defines the reliable usable upper frequency limit as 15%–20% of the natural resonant frequency. Operating beyond this range causes sharp sensitivity amplification, resonant overshoot, waveform distortion and invalid measured data.
Based on a 250 kHz natural resonant frequency, the reliable high‑frequency bandwidth of this sensor is approximately 37.5–50 kHz. This bandwidth fully covers the signal spectrum of most blast shock waves, water hammer transients and pipeline pressure pulsations.
Technical Note: Natural resonant frequency is not governed solely by the sensing wafer. Overall structural stiffness, housing geometry, diaphragm thickness and installation preload exert significant influence on the realized resonant frequency. Improper installation torque or insufficient stiffness of the mounting fixture will degrade the effective resonant frequency and distort acquired waveforms.
The Y1004E‑4 has a typical rise time of approximately 2 μs, defined as the interval required for the output signal to transition from 10% to 90% of its steady‑state amplitude under step pressure excitation.
Rise time is inversely correlated with natural resonant frequency: higher natural frequency yields shorter rise time and superior capability to resolve steep pressure fronts. The rising edge of blast overpressure commonly lasts only several microseconds. If the sensor rise time is substantially longer than the event timescale, the true pressure peak will be attenuated, producing an artificially low measured peak.
Important distinction: The specified rise time is an intrinsic parameter of the sensor. The overall response of the measurement chain is additionally constrained by cable capacitance, charge amplifier characteristics and data acquisition bandwidth. Even with a fast sensor, insufficient bandwidth of downstream instrumentation will roll off the shock front. Bandwidth matching across the full measurement chain is mandatory for blast pressure tests.
Housing Construction: The monolithic housing is fabricated from high‑strength stainless steel, balancing stiffness, shock resistance and vibration tolerance. Elevated stiffness improves mechanical robustness, raises the global natural frequency and suppresses parasitic structural resonance. The miniature sensor weighs merely 3 g. Its compact lightweight form enables installation in confined locations and minimizes flow field perturbation induced by mounting, suitable for flush mounting inside blast vessels and pipeline walls.
Electrical Interface & Power Supply: Equipped with a miniature L5 coaxial connector; an L5‑to‑BNC coaxial cable is provided as standard. It operates under constant current excitation: +18~28 VDC, 4–20 mA constant current source. It adopts ICP piezoelectric sensor architecture and outputs voltage signals. No standalone charge amplifier is required, simplifying field wiring and system commissioning.
Static Performance & Measurement Robustness: Measurement range 0–100 MPa, overload capacity 120% FS. Non‑linearity, hysteresis and repeatability are all < 1% FS. Overload margin is critical for shock wave testing. Blast overpressure is frequently accompanied by transient high‑amplitude pressure spikes. Sufficient overload capability protects the sensing element from transient overpressure damage and guarantees test reproducibility.
✅ Blast and detonation shock wave pressure measurement
✅ Pipeline water hammer and transient fluid impact pressure
✅ Pulsating pressure acquisition for internal combustion engines and pulsed machinery
✅ Fluid impact and gas explosion pressure tests
6. Summary of Key Technical Specifications
Piezoelectric sensors are designed for dynamic transient pressure measurement only. Do not deploy for long‑duration static pressure measurement.
The mounting base must possess adequate stiffness. Fastener torque shall strictly follow the datasheet specification. Elastic deformation of the base reduces the effective resonant frequency.
The bandwidth of the complete measurement chain (cable, ICP signal conditioner, data acquisition unit) must exceed the bandwidth of the target signal; otherwise the sensor’s high‑speed performance cannot be realized.
For blast test environments, install sealing gaskets for protection to prevent diaphragm damage from high‑temperature fragments.
Yangzhou Xiyuan Electronic Technology Co.,Ltd.
#77 Yeqiao Road, Liandong U Valley slender West Lake Innovation port, building 10-2,Hanjiang District, Yangzhou,Jiangsu,China
Mobie Phone: +86 180-5105-8377
Tel: +86 514-82885589
Whatsapp:+86 182-6066-6867
Email: sale1@yzxyt.com
Fax: +86 514-82885089